DSI/2025/005 - Profemur Cobalt Chrome Modular Neck Hip Replacements: Higher risk of revision surgery, metal-wear effects and component fracture.

Documents

Details

An MHRA investigation has found increased risks of wear and corrosion, including an increased occurrence of device fracture and revision surgery associated with cobalt chrome-containing Profemur modular neck hip stem components.

Patients implanted with affected Profemur devices are to be invited for a clinical review to inform them of the identified risks, assess their clinical presentation and discuss the need for continued follow-up.